Directions for use Digital MacroView 5
Cautions
A caution statement in this manual identifies a condition or
practice, which if not corrected or discontinued immediately,
could lead to equipment failure, equipment damage, or data loss.
Caution In order to obtain optimum performance and
ensure patient safety, thoroughly read this booklet.
Caution The Digital MacroView System is designed
expressly for use with an approved Welch Allyn 3.5v
power source. Damage to the system will occur if
alternate power sources are attached to the video head.
Caution Do not use the Digital MacroView if you
notice any signs of damage to the components of the
system. Contact Customer Service Department for
immediate assistance.
Caution Do not attempt to disinfect the Digital
MacroView System using glutaraldehyde products,
ethylene oxide gas, steam or any other liquid or gas
disinfectant.
Caution The signal output (USB Port) connector is
intended for connection of only devices complying with
IEC 60601-1, or other IEC standards (for example, IEC
60950), as appropriate to the device. Connecting
additional devices to the Digital MacroView System is
configuring a medical system and you are responsible
for verifying that the system complies with the
requirements of the system standard IEC 60601-1-1.
Caution This product complies with current required
standards for electromagnetic interference and should
not present problems to other equipment or be affected
by other devices. As a precaution, avoid using this
device in close proximity to other equipment.