Hach Polymetron 9582sc Basic User Manual

Tyyppi
Basic User Manual
DOC023.98.93067
Polymetron 9582sc
Dissolved Oxygen Analyzer
05/2018, Edition 5
Basic User Manual
Allgemeines Benutzerhandbuch
Manuale di base per l'utente
Manuel d'utilisation de base
Manual básico del usuario
Manual básico do utilizador
基本用户手册
Základní uživatelská příručka
Basisgebruikershandleiding
Grundlæggende brugervejledning
Allmän användarhandbok
Peruskäyttöohje
Alapvető felhasználói kézikönyv
Начальное руководство пользователя
Temel Kullanıcı Kılavuzu
English..............................................................................................................................3
Deutsch.......................................................................................................................... 20
Italiano............................................................................................................................ 39
Français......................................................................................................................... 58
Español.......................................................................................................................... 77
Português...................................................................................................................... 96
中文............................................................................................................................... 115
Čeština......................................................................................................................... 132
Nederlands................................................................................................................. 150
Dansk............................................................................................................................169
Svenska....................................................................................................................... 187
Suomi............................................................................................................................206
Magyar......................................................................................................................... 223
Русский........................................................................................................................242
Türkçe...........................................................................................................................261
2
Table of contents
Specifications on page 3 Operation on page 9
General information on page 4 Maintenance on page 13
Installation on page 5 Troubleshooting on page 16
Startup on page 9 Replacement parts and accessories on page 19
User navigation on page 9
Expanded manual version
For additional information, refer to the expanded version of this manual, which is available on the
manufacturer's website.
Specifications
Specifications are subject to change without notice.
Sensor
Specification Details
Sensor body Noryl
Sensor body with optional immersion Stainless steel 316L
Dimensions 144 x 144 x 181 mm (5.7 x 5.7 x 7.1 in.)
Weight 1.7 kg (3.75 lb)
Electrode Cathode: gold; Anode: silver
Membrane PFA
Membrane holder Noryl
Sample flow rate 4—10 liters/hour
Sample temperature -20—60 °C (-4—120 °F)
Sample pressure Atmospheric pressure
Relative humidity 0—90%
Certifications EN 61326-1: 2006; EN 61010-1: 2010
Sensor module
Specification Details
Measuring range 0—2 ppm
Sensitivity < 0.5 ppb
Repeatability ± 0.5 ppb or ± 2% of measurement whichever is the greater
Detection limit ≤ 1 ppb
Response time 1—40 ppb: < 30 seconds
Sample temperature measurement range 0—45 °C (32—113 °F)
English 3
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
W A R N I N G
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential
injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol indicates the presence of devices sensitive to Electro-static Discharge (ESD) and
indicates that care must be taken to prevent damage with the equipment.
This symbol, when noted on a product, indicates the instrument is connected to alternate current.
Electrical equipment marked with this symbol may not be disposed of in European domestic or
public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no
charge to the user.
4 English
Products marked with this symbol indicates that the product contains toxic or hazardous substances
or elements. The number inside the symbol indicates the environmental protection use period in
years.
Products marked with this symbol indicates that the product conforms to relevant South Korean
EMC standards.
EMC compliance statement (Korea)
Type of equipment Additional information
A 기기
( 업무용 방송통신기자재 )
기기는 업무용 (A ) 전자파적합기기로서 판매자
사용자는 점을 주의하시기 바라며, 가정외의 지역
에서 사용하는 것을 목적으로 합니다.
Class A equipment
(Industrial Broadcasting and Communication
Equipment)
This equipment meets Industrial (Class A) EMC
requirements. This equipment is for use in industrial
environments only.
Product overview
This single-channel analyzer is used for the measurement of dissolved oxygen in boiler feedwaters,
economizers, condensers and in general all thermal equipment using water as a heat transfer liquid.
Product components
Make sure that all components have been received. If any items are missing or damaged, contact the
manufacturer or a sales representative immediately.
Components:
Controller (contains instructions for controller installation and operation)
Sensor (contains instructions for sensor installation and operation)
Sensor cable
Flow chamber
Sensor maintenance kit (contains membranes and electrolyte)
Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Installation guidelines
Put the instrument in a location that has access for operation, service and calibration.
Make sure that there is good view of the display and controls.
Keep the instrument away from a heat source.
Keep the instrument away from vibrations.
Keep the sample tubing as short as possible to minimize the response time.
Make sure that there is no air in the sample supply line.
English
5
Assemble the sensor
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Electrolyte must be added to the sensor before use. Refer to the steps that follow and Figure 1.
1. Install the membrane on the sensor body by hand as far as it will go.
2. Use the syringe to add 5 mL of electrolyte to the sensor body. Make sure that there are no
impurities or bubbles in the electrolyte.
3. Gently push the electrode into the sensor body until there is resistance. Do not force the
electrode.
4. Install and tighten the holding nut by hand.
5. Install the gasket and filling screw to prevent electrolyte leakage or external contamination from
the sample. Carefully tighten the screw with a screwdriver without using excessive force.
Figure 1 Sensor assembly
Install the sensor in the process
N O T I C E
Keep the sensor in a vertical position with the membrane down during mounting and removal. Do not shake the
sensor so that oxygen does not contaminate the electrolyte.
Refer to the steps that follow and Figure 2 to mount the sensor in the flow chamber.
1. Connect the sensor cable to the sensor and tighten the connector by hand. Do not turn the cable
or the sensor.
2. Put the sensor into the flow chamber and tighten the sensor clamping nut by hand.
3. Connect the sample line to the sample input connector.
4. Connect the sample output to the sample output connector.
6
English
Figure 2 Sensor mounting
1 Sensor cable connector 5 Flow chamber
2 Sensor clamping nut 6 1/4" NPT sample input connector (not provided)
3 Assembled sensor 7 M4 screws (not provided)
4 Gasket 8 1/4" NPT sample output connector (not provided)
Install the controller
Refer to the controller documentation for mounting and wiring instructions.
Connect the sensor to the controller
W A R N I N G
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical
connections.
W A R N I N G
Electrocution Hazard. High voltage wiring for the controller is conducted behind the high voltage barrier in the
controller enclosure. The barrier must remain in place except when installing modules, or when a qualified
installation technician is wiring for power, relays or analog and network cards.
English 7
N O T I C E
Potential Instrument Damage. Delicate internal electronic components can be damaged by static
electricity, resulting in degraded performance or eventual failure.
Refer to the illustrated steps that follow and Table 1.
Be sure to connect all sensor ground/shield wires to the controller enclosure grounding screws.
Note: If the sensor cable is not long enough to reach the controller, an interconnect cable and junction box are
required to extend the distance.
Table 1 Sensor wiring
Terminal Signal Sensor wire
1 Temp + Black
2 Temp – Blue
3–4
5 Ground Green
6 Earth Yellow
7–8
9 Working electrode White
10 Counter electrode Red
11–12
8 English
Startup
Make sure that the flow rate and pressure do not exceed the values in Specifications on page 3.
1. Open the valve on the sample line to let sample flow through the analyzer.
2. Turn the knob on the flow meter to set the flow rate.
3. Examine the plumbing for leaks and stop any leaks if found.
4. Apply power to the controller.
5. Make the applicable menu selections when the controller starts.
User navigation
Refer to the controller documentation for keypad description and navigation information.
Operation
System configuration
Refer to the controller documentation for system configuration, general controller settings and setup
for outputs and communications.
English
9
Configure the sensor
Use the CONFIGURE menu to enter identification information for the sensor and to change options
for data handling and storage.
1. Push the menu key, select SENSOR SETUP>CONFIGURE.
2. Use the arrow keys to select an option and push enter. To enter numbers, characters or
punctuation, push and hold the up or down arrow keys. Push the right arrow key to advance to
the next space.
Option Description
EDIT NAME Changes the name that corresponds to the sensor on the top of the measure screen.
The name is limited to 16 characters in any combination of letters, numbers, spaces or
punctuation.
SENSOR S/N Allows the user to enter the serial number of the sensor, limited to 16 characters in any
combination of letters, numbers, spaces or punctuation.
MEAS UNITS Changes the measurement units—Select the unit from the list available.
PRESSURE UNITS Sets the units for atmospheric pressure—Select the unit from the list available.
TEMP UNITS Sets the temperature units to °C (default) or °F
FILTER Sets a time constant to increase signal stability. The time constant calculates the
average value during a specified time—0 (no effect, default) to 60 seconds (average of
signal value for 60 seconds). The filter increases the time for the sensor signal to
respond to actual changes in the process.
LOG SETUP Sets the time interval for data storage in the data log—5, 30 seconds, 1, 2, 5, 10,
15 (default), 30, 60 minutes.
RESET DEFAULTS Sets the configuration menu to the default settings. All sensor information is lost.
Calibrate the sensor
About sensor calibration
The sensor characteristics slowly shift over time and cause the sensor to lose accuracy. The sensor
must be calibrated regularly to maintain accuracy. The calibration frequency varies with the
application and is best determined by experience.
Temperature calibration
It is recommended to calibrate the temperature sensor once a year. Calibrate the temperature sensor
before calibrating the measurement sensor.
1. Put the sensor in a container of water that is at a known temperature. Under agitation, measure
the temperature of the water with an accurate thermometer or independent instrument.
2. Push the menu key and select SENSOR SETUP>CALIBRATE.
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select 1 PT TEMP CAL and push enter.
5. The raw temperature value is displayed. Push enter.
6. Enter the correct value if different from that displayed and push enter.
7. Push enter to confirm the calibration. The temperature offset is displayed.
Zero calibration
Because of the stability of the electrode, for most applications this calibration is not required but can
be used to define the unique zero point of the sensor.
1. Remove the sensor from the process and rinse in distilled water.
2. Push the menu key and select SENSOR SETUP>CALIBRATE.
10
English
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select ZERO CAL and push enter.
5. Select the option for the output signal during calibration:
Option Description
ACTIVE The instrument sends the current measured output value during the calibration procedure.
HOLD The sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
6. Place the clean sensor in a zero concentration solution, push enter.
7. Wait for up to one hour for the value to stabilize and push enter.
8. Review the calibration result:
PASS—the sensor is calibrated and the offset is displayed.
FAIL—the calibration is outside of accepted limits. Clean the sensor and retry with a fresh
reference solution. Refer to Troubleshooting on page 16 for more information.
9. If the calibration passed, push enter to continue.
10. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to Change calibration options on page 13.
11. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YES The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NO The sensor was calibrated previously with this controller.
12. Return the sensor to the process and push enter. The output signal returns to the active state
and the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Calibration in air
Air calibration is recommended for best accuracy and repeatability.
1. Remove the sensor from the process.
2. Push the menu key and select SENSOR SETUP>CALIBRATE.
3. If the pass code is enabled in the security menu for the controller, enter the pass code.
4. Select AIR CAL and push enter.
5. Select the option for the output signal during calibration:
Option Description
ACTIVE The instrument sends the current measured output value during the calibration procedure.
HOLD The sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
6. Using a precision certified barometer, measure the atmospheric pressure where the analyzer is
located. Use the arrow keys to enter this value if different from the value displayed and push
enter.
7. Moisturize the wadding in a calibration cap with a few drops of water. Place the sensor vertically
in the calibration cap with the membrane downwards. Tighten the calibration cap and push enter.
8. Wait for the value to stabilize and push enter.
English
11
9. Review the calibration result:
PASS—the sensor is calibrated and the calibration factor is displayed.
FAIL—the calibration is outside of accepted limits. Repeat the calibration with fresh reference
solutions. Refer to Troubleshooting on page 16 for more information.
10. If the calibration passed, push enter to continue.
11. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID.
12. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YES The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NO The sensor was calibrated previously with this controller.
13. Return the sensor to the process and push enter.
The output signal returns to the active state and the measured sample value is shown on the
measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Calibration with the process sample
The sensor can remain in the process sample.
1. Push the menu key and select SENSOR SETUP>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select SAMPLE CAL and push enter.
4. Select the option for the output signal during calibration:
Option Description
ACTIVE The instrument sends the current measured output value during the calibration procedure.
HOLD The sensor output value is held at the current measured value during the calibration procedure.
TRANSFER A preset output value is sent during calibration. Refer to the controller user manual to change
the preset value.
5. With the sensor in the process sample, push enter. The measured value is shown. Wait for the
value to stabilize and push enter.
6. With a certified secondary verification instrument measure the concentration value of the sample.
To avoid impurities in the sample take the measurement before the sample enters the flow
chamber. Use the arrow keys to enter this value if different from the value displayed and push
enter.
7. Review the calibration result:
PASS—the sensor is calibrated and the calibration factor is displayed.
FAIL—the calibration is outside of accepted limits. Clean the sensor and retry. Refer to
Troubleshooting on page 16 for more information.
8. If the calibration passed, push enter to continue.
9. If the option for operator ID is set to YES in the CAL OPTIONS menu, enter an operator ID. Refer
to Change calibration options on page 13.
10. On the NEW SENSOR screen, select whether the sensor is new:
Option Description
YES The sensor was not calibrated previously with this controller. The days of operation and previous
calibration curves for the sensor are reset.
NO The sensor was calibrated previously with this controller.
12 English
11. With the sensor still in the process push enter. The output signal returns to the active state and
the measured sample value is shown on the measure screen.
Note: If the output mode is set to hold or transfer, select the delay time when the outputs return to the active
state.
Exit calibration procedure
If the back key is pushed during a calibration, the user can exit the calibration.
1. Push the back key during a calibration. Three options are shown:
Option Description
QUIT CAL Stop the calibration. A new calibration must start from the beginning.
BACK TO CAL Return to the calibration.
LEAVE CAL Exit the calibration temporarily. Access to other menus is allowed. To return to the
calibration, push the menu key and select SENSOR SETUP.
2. Use the arrow keys to select one of the options and push enter.
Change calibration options
The user can set a calibration reminder or include an operator ID with calibration data from this
menu.
1. Push the menu key and select SENSOR SETUP>CALIBRATE.
2. If the pass code is enabled in the security menu for the controller, enter the pass code.
3. Select CAL OPTIONS and push enter.
4. Use the arrow keys to select an option and push enter.
Option Description
CAL REMINDER Sets a reminder for the next calibration in days, months or years—select the required
delay from the list.
OP ID ON CAL Includes an operator ID with calibration data—YES or NO (default). The ID is entered
during the calibration.
Standby function
The measuring range of the 9582 is up to a maximum of 2 ppm. If measurements exceed this value
for more than 2.5 minutes the standby function is enabled and no more measurements are taken. To
exit standby mode push the menu key and select SENSOR SETUP>EXIT STANDBY.
Maintenance
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Membrane replacement
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
English 13
N O T I C E
Always remove the electrolyte filling plug before removal of the electrode. Do not touch the membrane with the
hands. Do not use a worn membrane.
Remove the sensor
1. Turn off the sample supply.
2. Unscrew the sensor cable connector from the sensor. Do not turn the actual cable or the sensor
itself.
3. Unscrew the sensor clamping nut from the flow chamber and remove the sensor (Figure 3).
Figure 3 Remove the sensor
Replace the membrane
Refer to the steps that follow and Figure 4 to replace the sensor membrane.
1. Unscrew the filling screw and remove along with the gasket.
2. Unscrew the holding nut.
3. Carefully remove the electrode from the sensor body.
4. Empty any electrolyte left in the sensor body.
5. Unscrew the old membrane from the sensor body.
6. Install the new membrane. Refer to Assemble the sensor on page 6.
14
English
Figure 4 Replace the membrane
Electrode rejuvenation
After some months of operation (3 to 12 depending on sample oxygen concentration, plant shutdown
frequency, etc.), a dark coating of silver bromide (AgBr) may cover part of the silver anode. This
coating does not affect the measurement unless more than 90% of the surface is contaminated.
When changing the electrolyte and membrane, visually check the silver anode. If more than 2/3 of
the surface is coated with silver bromide then an electrode rejuvenation is required. To do this, very
softly polish the areas coated with the deposit using a soft abrasive (N° 400 to 600). After polishing,
rinse with demineralized water and wipe dry with a soft cloth. Some of the dark coloring may be left
to improve sensor stabilization time.
After cleaning, replace the sensor in the sample and leave for 30 minutes for the measurement to
stabilize. Once stable, the sensor must be calibrated.
English
15
Troubleshooting
General troubleshooting
Problem Probable cause Resolution
Excessive time for
stabilization or no
stabilization during
calibration in air.
The sample temperature is very different
from the ambient temperature, e.g. 6 °C
(43 °F) in water and 35 °C (95 °F) in air will
cause a measurement drift.
Do not wait till the sensor temperature
equals the external temperature (use
temperature compensation).
There is an electrolyte leak through the
membrane. The current is too high because
of an excessive penetration of oxygen.
Change the membrane.
There is pollution of the electrolyte due to a
loose filling screw.
Change the electrolyte. Check there is a
gasket in place and tighten the screw using
a screwdriver but without using excessive
force.
The electrode is not correctly mounted in
the sensor body causing an excessive gap
between the membrane and the cathode.
Tighten the electrode holding nut.
The membrane is incorrectly installed
causing a risk of electrolyte pollution.
Change the electrolyte and reinstall the
membrane onto the sensor body as far as it
will go finger tight.
Lack of humidity due to high temperature.
Use a calibration cap (see Replacement
parts and accessories on page 19).
There is water or humidity in the sensor
cable connector.
Dry the sensor cable connector inside and
out, and install finger tight.
The gold surface is scratched or damaged. Change the electrode.
Sludge or particles on the cathode.
Clean the cathode with a soft and
absorbent tissue. Rinse the membrane.
Cable or connections damaged when the
sensor has been removed.
Check the sensor connection to the
controller module. If correct check the
connections to the sensor cable connector.
The sensor is incorrectly positioned. The
electrolyte has leaked and air bubbles have
entered the cathode.
Put the sensor in the correct position, head
down.
No significant
current increase
when the sensor is
in the air for the
calibration.
There is pollution of the electrolyte due to a
loose filling screw.
Change the electrolyte. Check there is a
gasket in place and tighten the screw using
a screwdriver but without using excessive
force. Check the sensor is not damaged.
There is pollution of the electrolyte due to a
leaking membrane.
Change the electrolyte and the membrane.
The membrane is torn. Change the membrane.
The electrode is not correctly mounted in
the sensor body causing an excessive gap
between the membrane and the cathode.
Tighten the electrode holding nut.
Cable or connections damaged when the
sensor has been removed.
Check the connection to the controller
module. If correct check the connections to
the sensor cable connector.
The membrane is worn. Change the membrane.
There is a dark green deposit of silver
bromide (AgBr) on the silver tube.
Polish the tube with a soft abrasive (N°
400 to 600) and replace the membrane.
16 English
Problem Probable cause Resolution
Important instability
in measuring mode.
There is water or humidity in the sensor
cable connector.
Dry the sensor cable connector inside and
out, and install finger tight.
Incorrect connection.
Check the sensor connection to the
controller module.
There are bubbles close to the cathode.
Refill with electrolyte and check there are
no bubbles at the bottom of the sensor
body.
The sensor has been violently shaken.
Check the fixation and stability of the
sensor.
Electromagnetic interferences close to the
sensor or controller cable.
Find a better place for the cable and check
the EMC levels.
Temporary interference with other gases. Mainly with H
2
S.
Flow rate too low (4 mL/h minimum). Increase sample flow.
Sludge from a heterogeneous sample has
damaged the membrane.
Install a deflector or change the sensor
location.
Pressure variation in the line.
Make sure the sensor is used at
atmospheric pressure.
Lack of accuracy.
The permeability of the membrane has
changed (dirt deposits).
Calibrate the analyzer and check if the
concentration is back to normal.
Electrolyte pollution.
Check the screw-in parts (membrane, filling
screw) and change the electrolyte and the
membrane.
Electrolyte leaking.
Check the screw-in parts (membrane, filling
screw) and change the electrolyte and the
membrane.
Interferences, mainly with H
2
S.
If the level of H
2
S (or other pollutant) is
stable, take its concentration into account to
determine the dissolved O
2
concentration.
Error during calibration or incorrect
calibration.
Calibrate again to check the parameters. If
the error persists check the calibration
current (too high, too low or unstable) and
the concentration in air. Refer to the
problems described above.
There are bubbles close to the cathode.
Refill with electrolyte and check there are
no bubbles at the bottom of the sensor
body.
Flow rate too low (4 mL/h minimum). Increase sample flow.
The sample temperature or pressure is out
of specification.
Change the sensor location or modify the
sample so that it is within specification.
Sludge or particles on the cathode.
Clean the cathode with a soft and
absorbent tissue. Change the membrane.
The sensor current
is null during
measurement.
There is no electrolyte in the sensor
(leakage).
Check the screw-in parts (membrane, filling
screw) and change the electrolyte.
Sensor cable is disconnected or poorly
connected.
Check cable connections on the module
and the sensor connector in the controller.
English 17
Problem Probable cause Resolution
The sensor current
is negative.
Connection problem to the anode circuit
(loose contact).
Check the connection to the controller
module. If correct check the connections to
the sensor cable connector.
There is a dark green deposit of silver
bromide (AgBr) on the surface of the silver
tube.
Polish the tube with a soft abrasive (N°
400 to 600).
The sample
temperature is out of
specification.
There may be a short-circuit on the
temperature connection.
Check the connection to the controller
module. If correct check the connections to
the sensor cable connector.
The displayed
reading is not
numeric.
The measured value is < 0 ppb if negative
signs are displayed.
Check the zero calibration.
The measured value is > 10,000 ppb Change the display unit.
The standby function is enabled because
the measured value is > 2 ppm.
Exit the standby function from the menu
option.
Sensor diagnostic and test menu
The sensor diagnostic and test menu shows current and historical information about the instrument.
Refer to Table 2. To access the sensor diagnostic and test menu, push the MENU key and select
Sensor Setup, [Select Sensor], DIAG/TEST.
Table 2 Sensor DIAG/TEST menu
Option Description
MODULE INFORMATION Shows the version and the serial number for the sensor module.
SENSOR INFORMATION Shows the name and serial number that was entered by the user.
CAL DAYS Shows the number of days since the last calibration.
CAL HISTORY Shows a list of the calibrations and the details for each calibration.
RESET CAL HISTORY Service use only. Resets the calibration history for the sensor. All previous calibration
data is lost.
SENSOR SIGNALS Shows the current reading in mV and the temperature.
MEMBRANE DAYS Shows the number of days that the sensor has been in operation.
RESET MEMBRANE Resets the number of days that the sensor has been in operation.
Warning list
A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in Table 3.
Table 3 Warning list for dissolved oxygen sensors
Warning Description Resolution
DO TOO HIGH The measured value is > 40 ppm Make sure that the DO level in the
process water is within the
operating limits of the sensor.
Calibrate or replace the sensor.
DO TOO LOW The measured value is < 0 ppb Calibrate or replace the sensor.
18 English
Table 3 Warning list for dissolved oxygen sensors (continued)
Warning Description Resolution
TEMP TOO HIGH The measured temperature is >
50 °C
Reduce sample temperature.
TEMP TOO LOW The measured temperature is <
0 °C
Increase sample temperature.
CURRENT TOO HIGH The measured current > 200 µA Make sure that the DO level in the
process water is within the
operating limits of the sensor.
Calibrate or replace the sensor.
CURRENT TOO LOW The measured current < -0.5 mA Calibrate or replace the sensor.
CAL OVERDUE The Cal Reminder time has expired Calibrate the sensor.
REPLACE SENSOR The sensor has been in operation >
365 days
Replace the sensor cartridge and
calibrate the sensor. If the
calibration result is pass, reset the
membrane days in the DIAG/TEST
menu.
NOT CALIBRATED The sensor has not been calibrated Calibrate the sensor.
CAL IN PROGRESS A calibration was started but not
completed
Return to calibration.
Replacement parts and accessories
Refer to the replacement parts and accessories section of the controller documentation for controller
parts and accessories.
Note: Product and article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts and accessories
Description Item no.
Box of 4 pre-mounted membranes 09185=A=3500
Calibration cap 09182=A=1200
Electrolyte filling screw 09078=C=1020
Electrolyte filling washer 09078=C=1030
Oxygen electrode without sensor body ppb 09182=A=1000
Oxygen sensor body ppb 09078=C=1010
Reference electrolyte 25 mL 09181=A=3600
Stainless steel flow cell 09078=A=2000
Syringe 460150,21951
English 19
Inhaltsverzeichnis
Spezifikationen auf Seite 20 Betrieb auf Seite 27
Allgemeine Informationen auf Seite 21 Wartung auf Seite 31
Installation auf Seite 22 Fehlerbehebung auf Seite 34
Inbetriebnahme auf Seite 27 Ersatzteile und Zubehör auf Seite 38
Benutzernavigation auf Seite 27
Erweiterte Version des Handbuchs
Zusätzliche Informationen finden Sie in der ausführlichen Version dieser Bedienungsanleitung auf
der Website des Herstellers.
Spezifikationen
Die Spezifikationen können ohne Vorankündigung Änderungen unterliegen.
Sensor
Spezifikation Details
Sensorgehäuse Noryl
Sensorgehäuse ohne optionale Immersion Edelstahl 316L
Abmessungen 144 x 144 x 181 mm (5,7 x 5,7 x 7,1 Zoll)
Gewicht 1,7 kg / 3,75 lb
Elektrode Kathode: gold; Anode: silber
Membran PFA
Membranhalterung Noryl
Probenflussrate 4—10 l/h
Probentemperatur -20—60 °C (-4—120 °F)
Probendruck Luftdruck
Relative Feuchtigkeit 0—90%
Zertifikationen EN 61326-1: 2006; EN 61010-1: 2010
Sensormodule
Spezifikation Details
Messbereich 0—2 ppm
Empfindlichkeit < 0,5 ppb
Wiederholbarkeit ± 0,5 ppb oder ± 2% der Anzeige, je nachdem, welcher Wert größer ist
Erfassungsgrenze < 1 ppb
Reaktionszeit 1—40 ppb: < 30 Sekunden
Temperaturmessbereich Probe 0—45°C (32—113°F)
20 Deutsch
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280

Hach Polymetron 9582sc Basic User Manual

Tyyppi
Basic User Manual